2
HOME PRODUCTS SH-MXT™-5

SH-MXT™-5

Product overview
SH-MXT™-5 Structure

- Low-polarity phase: Crossbond™ 5% diphenyl / 95% dimethyl polysiloxane
- General-purpose columns for drugs, solvent impurities, pesticides, hydrocarbons, PCB congeners (e.g., Aroclor mixes), essential oils, and semivolatiles.
- Equivalent to USP G27 and G36 phases.
- 4.5" standard coil diameter.
- Similar phases: DB-PS5, VF-5ht UltiMetal

 

Application

Phenols by EPA 604

 
Peaks
1. Phenol
2. 2-Chlorophenol
3. 2-Nitrophenol
4. 2,4-Dimethylphenol
5. 2,4-Dichlorophenol
Peaks  
6. 4-Chloro-3-methylphenol
7. 2,4,6-Trichlorophenol
8. 2,4-Dinitrophenol
9. 4-Nitrophenol
10. 2-Methyl-4,6-dinitrophenol
11. Pentachlorophenol
 
Column MXT-5, 30 m, 0.28 mm ID, 0.25 µm (cat.# 70224)
Sample 604 Phenols Calibration Mix (cat.# 31029)
Conc.: 25 ng/µL per component
Injection
Inj. Vol.: 1.0 µL splitless
Inj. Temp.: 280 °C
Oven
Oven Temp.: 40 °C to 250 °C at 10 °C/min
Carrier Gas H2, constant pressure
Linear Velocity: 50 cm/sec @ 40 °C
Detector: FID @ 300 °C
Notes: FID sensitivity: 2.56 x 10-10 AFS