SH-Rtx™-VMS

Product overview
- Dedicated columns for analyzing volatile organic pollutants by GCMS including methods TO-15, TMS, and EPA 8260.
- Complete separation of U.S. EPA Method 8260 compounds in less than 10 minutes.

 

Application

Volatile Organics US EPA CLP 04.1

Peaks
1. Dichlorodifluoromethane
2. Chloromethane
3. Vinyl chloride
4. Bromomethane
5. Chloroethane
6. Trichlorofluoromethane
7. 1,1-Dichloroethene
8. Carbon disulfide
9. 1,1,2-Trichloro-1,2,2-Trifluoroethane
10. Methylene Chloride
11. Acetone
12. trans-1,2-Dichloroethene
13. Methyl acetate
14. Methyl tert-butyl ether
15. 1,1-Dichloroethane
16. cis-1,2-dichoroethane
17. Cyclohexane
18. Bromochloromethane (IS)
Peaks  
19. Chloroform
20. Carbon Tetrachloride
21. 1,1,1-Trichloroethane
22. 2-Butanone
23. Benzene
24. 1,2-dichloroethane-d4 (SS)
25. 1,2-Dichloroethane
26. Methylcyclohexane
27. Trichloroethene
28. 1,4-Difluorobenzene (IS)
29. 1,2-Dichloropropane
30. Bromodichloromethane
31. cis-1,3-Dichloropropene
32. toluene d8 (SS)
33. Toluene
34. Tetrachloroethane
35. 4-Methyl-2-pentanone
36. trans-1,3-dichloropropane
Peaks  
37. 1,1,2-Trichloroethane
38. Dibromochloromethane
39. 1,2-Dibromoethane
40. 2-Hexanone
41. chlorobenzene d5 (IS)
42. Chlorobenzene
43. Ethylbenzene
44. m-Xylene
45. p-Xylene
46. o-Xylene
47. Styrene
48. Bromoform
49. Isopropylbenzene
50. 4-Bromofluorobenzene (SS)
51. 1,1,2,2-Tetrachloroethane
52. 1,3-Dichlorobenzene
53. 1,4-Dichlorobenzene
54. 1,2-Dichlorobenzene
55. 1,2-Dibromo-3-Chloropropane
56. 1,2,4-Trichlorobenzene

 
Column:   SH-Rtx™-VMS, 30 m, 0.25 mmID, 1.40 μm (P/N: 227-36268-01)
Purge and Trap:   Trap: #10 (TenaxR/silica gel/carbon molecular sieve)
  Sample Temp: ambient
  Purge: 11 min at 40 mL/min
  Desorb preheat: 185 °C
  Desorb: 0.5 min at 190 °C
  Desorb flow rate: 35.0 mL/min
  Bake: 8 min at 210 °C
  Interface: split injector
  Transfer Line Temp: 150 °C
Inj.:   Split (split ratio: 35:1)
Inj. Temp:   200 °C
Carrier Gas:   He, linear velocity 34 cm/sec., 40 °C, constant flow
Oven Temp:   40 °C (hold 4 min) to 90 °C at 16 °C/min to 220 °C at 32 °C/min (hold 5 min)
Detector:   MS
  Transfer Line Temp: 150 °C
  Scan Range: 35-300 amu.
  Ionization: EI